Dwyer acting as series gfm2 gas mass flow meter
Welcome: KUNAG Fluid Control System (Shanghai) Co., Ltd.
Language: Chinese ∷  English
Search

Dwyer

  • Series GFM2 Gas Mass Flow Meter dwyer
Series GFM2 Gas Mass Flow Meter dwyer

Series GFM2 Gas Mass Flow Meter dwyer

  • dwyer product sales product selection product quotation technical support service
  • American Dewell pressure difference meter flowmeter differential pressure gauge differential pressure switch differential pressure transmitter
  • Products are widely used in HVAC, food, pharmaceutical, bioengineering, intelligent building, power plant, chemical industry, water treatment, environmental monitoring, industrial manufacturing, etc.
  • Product description: Dwyer Instruments, Inc. of the United States (Dwyer Instruments, Inc.) is a company that has many precision instrumentation products for the measurement, transmission and control of differential press
  • INQUIRY
Series GFM2
Gas Mass Flow Meter is ideal for measurement of flow rates in a variety of gases. Two relay outputs and programmable totalizer that indicates total gas quantity.
±1% FS, Programmable Relays
The SERIES GFM2 Gas Mass Flow Meter is an ideal choice for the measurement of flow rates of a wide variety of gases. Unit can be calibrated for a variety of gases with user selectable 0 to 5 VDC or 4 to 20 mA and two relay outputs and programmable totalizer that indicates total gas quantity.
California Residents:  Click Here for Proposition 65 WARNING.
Service: Clean gases compatible with wetted parts.
Wetted Materials:
GFM2-X-X-A: Anodized aluminum, brass, 316 SS fluoroelastomer O-rings;
GFM2-X-X-S: 316 SS, and fluoroelastomer O-rings; Buna-N, EPR and PTFE O-rings optional.
Accuracy: ±1% FS.
Repeatability: ±0.25% FS.
Response Time: 2 seconds to within ±2% of actual flow.
Output Signal: Linear 0 to 5 VDC (3000 Ω min. load impedance) and 4 to 20 mA (500 Ω max. loop resistance).
Relay Rating: 1 A @ 24 VDC.
Max. Particulate Size: 5 microns.
Temperature Limits: 32 to 122°F (0 to 50°C).
Power Supply: 11 to 26 VDC.
Process Connections: 1/8" compression fitting for flow rates ≤10 L/min; 1/4" for ≤50 L/min; 3/8" for ≤100 L/min.
Display: 2 x 16 character LCD (optional).
Pressure Limits: 500 psig (34.5 bar).
Leak Integrity: 1 x 10-9 smL/sec of helium.
Weight: 1.05 lb (0.48 kg).

CATEGORIES

CONTACT US

Contact: peng gong

Phone: 18918462396

Tel: 021-59718851

Email: kunage@126.com

Add: Lane 819, Qinghewan Road, Qingpu Industrial Park, Qingpu District, Shanghai

Scan the qr codeClose