Welcome: KUNAG Fluid Control System (Shanghai) Co., Ltd.
Language: Chinese ∷  English
Search

Dwyer

  • 607 series differential pressure transmitter dwyer
607 series differential pressure transmitter dwyer

607 series differential pressure transmitter dwyer

  • dwyer product sales product selection product quotation technical support service
  • American Dewell pressure difference meter flowmeter differential pressure gauge differential pressure switch differential pressure transmitter
  • Products are widely used in HVAC, food, pharmaceutical, bioengineering, intelligent building, power plant, chemical industry, water treatment, environmental monitoring, industrial manufacturing, etc.
  • Product description: Dwyer Instruments, Inc. of the United States (Dwyer Instruments, Inc.) is a company that has many precision instrumentation products for the measurement, transmission and control of differential press
  • INQUIRY
607 series micro differential pressure transmitter
product manual:
Accuracy is ± 0.25% or ± 0.5%, ranging from 0-0.1 inch water column
607 series micro differential pressure transmitter can measure very small differential pressure, and has excellent stability and reliability, accuracy is ± 0.25% or ± 0.5%, and the measurement range is from 0-0.1 to 0-25 inches of water column, Wide range of uses. Ultra-thin glass fiber-coated silicon diaphragm can withstand vibration or vibration and eliminate drift. Each transmitter comes with a NIST traceability certificate. NEMA-2 protection grade, sturdy stainless steel housing protects it from moisture and dust. It can be used with air and other compatible gases.
Specifications:
Service: Air and non-conductive, non-corrosive gases.
Wetted Materials: Consult factory.
Accuracy: ± 0.5% or ± 0.25% FS.
Stability: ± 0.5% FS / yr.
Temperature Limits: -20 to 160 ° F (-29 to 71 ° C), 10 to 95% RH.
Pressure Limits: 10 psig (0.69 bar).
Compensated Temperature Range: 35 to 135 ° F (2 to 57 ° C).
Thermal Effects: ± 0.015% F.S./°F (zero and span).
Power Requirements: 12 to 36 VDC.
Output Signal: 4 to 20 mA DC, 2-wire.
Zero and Span Adjustments: Externally accessible potentiometers, non-interactive, ± 10% F.S. adjustment.
Response Time: 250 ms max.
Loop Resistance: 0 to 1045 ohms Vmin = 12V + [(. 22A) (RL)].
Current Consumption: 3.6 mA (min).
Electrical Connections: Screw terminals.
Process Connections: Barbed stainless steel for 3/16 "I.D. tubing.
Housing: 300 Series stainless steel (NEMA 2).
Weight: 1.04 lb (472 g).
Agency Approvals: CE.

CATEGORIES

CONTACT US

Contact: peng gong

Phone: 18918462396

Tel: 021-59718851

Email: kunage@126.com

Add: Lane 819, Qinghewan Road, Qingpu Industrial Park, Qingpu District, Shanghai

Scan the qr codeClose