Winding inductance F0 test method introduction
Winding inductance F0 test method 1. Test instrument Agilent E4991 or equivalent type of instrument; 2. Return the instrument to zero proofreading; 3. Place the sample to be tested on the instrument fixture; 4. Select the "Impedance" test option; 5. Slowly adjust the frequency adjustment knob to change the inductance test frequency (generally from 0HZ to 5GHz); 6. Investigate the impedance-frequency display curve appearing on the instrument screen, as shown in the following figure: Inductance_schematic 7. Find the peak point with the highest impedance (Impedance) on the picture; 8. The frequency point corresponding to the impedance peak point in the longitudinal direction is the self-resonant frequency point F0 of the inductor;