Electromagnetic flowmeter correction K coefficient formula introduction

The correction formula is as follows: K is new = Q mark / Q is × K,
The flow rate of the electromagnetic flowmeter is set to Q, and the flow of the electromagnetic flowmeter is Q. When the water meter Q is not equal to the Q of the inlet meter, we must correct the water meter and the sensor K factor.
Put K to be new, re-enter the converter, and observe the flow values of the inlet and outlet tables again to make them equal. The larger the K factor, the larger the flow rate.