Cause Analysis and Case Analysis of Electromagnetic Flowmeter Error
The electromagnetic flow
meter has many advantages and is widely used. However, if it is selected, installed and used improperly, it will cause an increase in error, unstable indication, and even damage to the body. This paper elaborates on the causes of electromagnetic flow
meter error, summarizes the experience and lessons of many years, and concludes that the causes of electromagnetic flow
meter error mainly have the following aspects.
1. The liquid in the electromagnetic flow
meter tube is not full. Due to insufficient back pressure or poor installation of the flow
sensor, the liquid in the measuring tube is not filled, and the fault phenomenon is different due to the degree of imperfection and the flow condition. If a small amount of gas is stratified or wavy in the water pipe, the fault phenomenon shows an increase in error, that is, the flow measurement value does not match the actual value; if the flow is a bubble flow or a plug flow, the fault phenomenon does not match the actual value. In addition, the output sloshing occurs due to the instantaneous covering of the electrode surface by the gas phase; if the gas flowing through the cross-sectional area of the horizontal pipe is increased, that is, the liquid is not full, the output sloshing may occur if the liquid is not full. The situation is so serious that the liquid level is below the electrode and the output is full.
Example 1 A shipyard has a SKLD series DN80mm electromagnetic flow
meter to measure the water flow. The operator reports that the flow rate is zero when the valve is closed, and the output reaches the full scale value. On-site inspection found that there was only a short tube downstream of the
sensor, the water was directly discharged into the atmosphere, and the shut-off valve was installed upstream of the
sensor. After the valve was closed, the water in the
sensor measuring tube was completely drained. By modifying the valve to position 2, the fault is solved.
The cause of such failures is often encountered in the case of the after-sales service of the manufacturer, which is an engineering design error.
Second, the liquid contains solids containing solids such as powder, granules or fibers, which may cause failures; 1 slurry noise; 2 electrode surface contamination; 3 conductive deposition layer or insulating deposition layer covering the electrode or lining; 4 lining is worn Or covered by sediment, the flow cross-sectional area is reduced. If the electromagnetic flow
sensor is used to measure the conductive lining of the insulating lining of the tube, the flow signal will be short-circuited to disable the
meter. Since the conductive material is gradually deposited, this type of fault usually does not appear in the commissioning period, but it does not appear until after a period of operation.
Example 2 Short-Circuit Effect of Conductive Deposit Layer. If the electromagnetic flow
sensor is used to measure the conductive lining of the insulating lining of the tube, the flow signal will be short-circuited to disable the
meter. Since the conductive material is gradually deposited, this type of fault usually does not appear in the commissioning period, but it does not appear until after a period of operation.
On the electrolysis cutting process test device of a diesel engine tool shop, the SKLD series DN80mm
meter is used to measure and control the saturated salt electrolyte flow rate to obtain the best cutting efficiency. At first, the
meter was operating normally. After 2 months of intermittent use, the flow display value was getting smaller and smaller until the flow signal was close to zero. On-site inspection revealed that a layer of yellow rust was deposited on the surface of the insulation layer, and the instrument was operating normally after wiping and cleaning. The yellow rust layer is caused by the deposition of a large amount of iron oxide in the electrolyte.
This example is a fault in the running period. Although it is not a common fault, if the ferrous metal pipeline is seriously corroded, the rust layer will be deposited. Anything that starts to function normally and the traffic is getting smaller and smaller over time should be analyzed for the possibility of such failures.
Third, liquids that may crystallize, electromagnetic flow
meters should be used with caution. Some easily crystallized chemical materials can be measured normally under normal temperature conditions. Because the conduits for transporting fluids have good heat insulation, they will not crystallize during heat preservation. However, the measuring tube of the electromagnetic flow
sensor is difficult to perform heat tracing. Therefore, when the fluid flows through the measuring tube, it is easy to cause a solid layer on the inner wall due to the temperature drop. Since the flow
meter measurement using other principles also has crystallization problems, in the absence of other better methods, a "ring" electromagnetic flow
sensor with a very short measuring tube length can be selected, and the upstream pipeline of the flow
meter can be used. Heat insulation is used to strengthen it. In the pipe connection method, it is convenient to disassemble and install the flow
sensor, and it is convenient to remove the maintenance once it is crystallized.
It is not uncommon for the example 3 to cause the electromagnetic flow
meter to malfunction due to liquid crystallization. For example, a smelter in Hunan installed a batch of electromagnetic flow
meters to measure the flow rate of the solution. Because the measuring tube of the electromagnetic flow
sensor is difficult to carry out heat insulation, a layer of crystals is formed on the inner wall and the electrode after a few weeks, causing the internal resistance of the signal source to become Very large, the instrument indicates that the value is abnormal. Because of the large dia
meter of these electromagnetic flow
meters, frequent washing and unloading is unbearable, so the end is still to use the open channel flow
meter.
Fourth, the electrode and grounding ring material selection problem caused by the material and the measured medium does not match the failure of the electromagnetic flow
meter and the medium contact parts have electrodes and grounding ring, mismatched in addition to corrosion resistance, as long as the electrode Surface effect. The surface effects should be: 1 chemical reaction (surface formation of blunt film, etc.); 2 electrochemical and polarization phenomena (potential generation); 3 catalyst action (molar formation on the electrode surface, etc.). Grounding rings also have these effects, but the impact is less.
Example 4 A chemical (smelting) plant in Shanghai used more than 20 Hastelloy B-electrode electromagnetic flow
meters to measure a higher concentration of hydrochloric acid solution, and the output signal was unstable. On-site inspection confirms that the instrument is normal and excludes other causes of interference that may cause output sloshing. However, it worked well in many other users using a Hastelloy B electrode
meter to measure hydrochloric acid. In the analysis of whether the cause of the failure is caused by the difference in hydrochloric acid concentration, the experience of the influence of the concentration of hydrochloric acid on the surface effect of the electrode should not be judged. For this reason, the instrument manufacturer and the user unit use the on-site conditions of the chemical plant to perform a real-flow test to change the concentration of hydrochloric acid. The concentration of hydrochloric acid gradually increases. When the concentration is low, the output of the
meter is stable. When the concentration increases to 15% to 20%, the output of the
meter begins to shake. When the concentration reaches 25%, the output sloshing amount is as high as 20%. After switching to the 钽 electrode electromagnetic flow
meter, it runs normally.
V. Problems caused by the liquid
conductivity exceeding the allowable range If the liquid
conductivity is close to the lower limit, there may be sloshing. Because the lower limit specified by the manufacturer's instrument specification is the lowest value that can be measured under various conditions of good use, and the actual conditions are not ideal, so low-grade distilled water or deionized water is encountered many times. Its
conductivity is close to the lower limit 5 specified by the electromagnetic flow
meter specification, and output sloshing occurs. It is generally considered that the lower limit of the
conductivity that can be stably measured is one to two orders of magnitude higher. The
conductivity of the liquid can be found in the relevant manual. If the data is not available, the sample can be measured with a
conductivity meter. However, there are cases where sampling from the pipeline is considered to be available in the laboratory and the actual electromagnetic flow
meter is not working. This is because the liquid at the time of measuring
conductivity is different from the liquid in the pipeline. For example, the liquid has absorbed CO2 or NO in the atmosphere to form carbonic acid or nitric acid, and the electrical
conductivity is increased.
For noise slurry produced by particles or fiber liquids, the method of increasing the excitation frequency can effectively improve the output sloshing. Some frequency-adjustable IFM3080F DN300 electromagnetic flow
meters measure 3.5% corrugated cardboard slurry and measure the instantaneous flow sloshing at different excitation frequencies in the field. When the frequency is lower, it is 50.32Hz, the shaking is as high as 10.7%; the frequency is increased to 50/2Hz, and the shaking is reduced to 1.9%, the effect is very obvious.